DaqLink Internal sensors
Internal Temperature Type: DBSA710: PT-100 DBSA720: Digital Range: -20 to 50 ºC Resolution: 0.1 ºC Accuracy: DBSA710: ±0.3 ˚C DBSA720: ±0.5 ˚C
Internal Humidity (for Logger PN: DBSA720) Range: 5 to 95 % Resolution: 0.5% Accuracy: ±3 % at 10 to 90 % ±5 % at 5 to 10 % and 90 to 95 %
DaqLink Inputs - DBSA710 and DBSA720 Input Types
4 to 20 mA Range: 4 to 20 mA Resolution: 4.76 μA Accuracy: ±0.5 % Loop impedance: 21 Ω Maximum load: 30 mA, 5.2 V
0 to 50 mV Range: 0 to 50 mV Resolution: 3 μV Accuracy: ±0.5 % Input impedance: 25 MΩ Maximum voltage: 5.2 V
0 to 1 V Range: 0 to 1 V Resolution: 200 μV Accuracy: ±0.5 % Input impedance: 25 MΩ Maximum voltage: 5.2 V
Temperature PT-100 (2-wires) Range: -200 to 400 ˚C Resolution: 0.1 ˚C Accuracy: -200 to -60 ˚C ±0.5 % -60 to 60 ˚C ±0.3 ˚C 60 to 400 ˚C ±0.5 %
Temperature Thermocouple J Range: -200 to 1,000 ˚C Resolution: 0.1 ˚C Accuracy: -200 to -60 ˚C ±0.5 % -60 to 60 ˚C ±0.5 ˚C 60 to 1,000 ˚C ±0.5 % Cold junction compensation error: ±0.3 ºC
Temperature Thermocouple K Range: -200 to 1,000 ºC Resolution: 0.1 ˚C Accuracy: -200 to -60 ˚C ±0.5 % -60 to 60 ˚C ±0.5 ˚C 60 to 1,000 ˚C ±0.5 % Cold junction compensation error: ±0.3 ˚C
Temperature Thermocouple T Range: -200 to 400 ˚C Resolution: 0.1 ˚C Accuracy: -200 to -60 ˚C ±0.5 % -60 to 60 ˚C ±0.5 ˚C 60 to 400 ˚C ±0.5 % Cold junction compensation error: ±0.3 ˚C
Frequency (input 4 only) Zero crossing detector Range: 20 Hz to 4 KHz Input signal: 0 to 4,000 Hz Input impedance: 470 Ω
Pulse Counter (input 4 only) Zero crossing detector Range: 1 to 65,536 counts Resolution: 1 count Frequency range: 0 to 4,000 Hz Input signal: 0 to 5 V Input impedance: 470 Ω
Contact Range: Open/Closed
Internal Temperature Type: DBSA710: PT-100 DBSA720: Digital Range: -20 to 50 ˚C Resolution: 0.1 ˚C Accuracy: DBSA710: ±0.3 ˚C DBSA720: ±0.5 ˚C
Internal Humidity (for DBSA720) Range: 5 to 95 % Resolution: 0.5 % Accuracy: ±3 % at 10 to 90 % ±5 % at 5 to 10 % and 90 to 95 %